Recall Z-0777-2013
Winco Mfg., LLC · initiated September 24, 2012
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PREMIER LIFECARE RECLINER 5400 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO AUGUST 20, 2012 A multi position recliner with a steel frame used primarily in clinics, hospitals, and treatment centers for recovery, dialysis, oncology, examination, infusion.
Codes / lots: Model 5400. Serial numbers: 540100017, 540100020, 540100063, 540100160, 540100161, 540100162, 540100062, 540100103, 540100104, 540100105, and 540100106.
Reason for recall
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
FDA-determined cause: Component design/selection
Action taken
Winco sent a Chair Field Correction Notification dated September 24, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The consignees were instructed to forward the "Field Correction Form - REPAIR" form back to Winco and the firm would send a Winco Representative to the consignees' facility to correct the chairs in the field. Customers with questions were instructed to contact Winco at 800-237-3377 or email customerservice@wincomfg.com. For questions regarding this recall call 352-854-2929, ext 110.
Details
- Recall number
- Z-0777-2013
- Initiated
- September 24, 2012
- Posted by FDA
- February 4, 2013
- Terminated
- May 23, 2013
- Product code
- INM – Chair, With Casters
- Quantity
- 95 units (all models)
- Distribution
- Nationwide Distribution including CA, FL, IL, MA, MD, MN, NM, NY, OH, OK, and TX.
- Recalling firm
- Winco Mfg., LLC, Ocala, FL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0778-2013 Class II | PREMIER CARE RECLINER 5570/5574 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO AUGUST…The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire… | 09/24/2012Terminated |
| Z-0779-2013 Class II | Cozy Comfort Premier Recliner 5580 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO…The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire… | 09/24/2012Terminated |
| Z-0780-2013 Class II | Designer Care Cliner 6700/6704/6710 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO…The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire… | 09/24/2012Terminated |