Recall Z-0786-2016

Bard Peripheral Vascular Inc · initiated January 5, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Bard Distaflo Vascular Bypass Graft; Product Code: DF8006SC; 6 mm x 80 cm. Intended for bypass or reconstruction of peripheral arterial blood vessels.

Codes / lots: Product Code: DF8006SC Lot Number: VTZE1783

Reason for recall

Size mislabeling; the graft is a 6 mm diameter graft when the actual packaged product is an 8 mm graft.

FDA-determined cause: Error in labeling

Action taken

Customer notification letters were sent on 01/05/2016. Customers were asked to not use the affected product, or further distribute it. Customers were asked to check all inventory locations for the affected product code, lot number combination listed. If customers further distributed any of the product they are requested to contact that location, advise them of the recall, and forward the listed instructions. The letter states to remove all identified product from the shelves. If no affected product is going to be returned the firm states to complete and return the attached form stating so. If customers have product it is requested that they also complete the attached form, and state the quantities and lot numbers in stock. The letter requests that customers call the firm's Recall Coordinator, Raye Seisinger, at 1-800-321-4254, option #2 ext. 2501 (M-F 6am to 3pm MST) or email raye.seisinger@crbard.com. Once all information has been verified the customer will be issued a Return Authorization or Consignment Recall Number to facilitate the expedient return of the product. The firm will provide replace product for returned product. Fax the attached form to the firm at 1-800-994-6772. If the customer cannot fax the firm they request that the Recall Coordinator be contact and provided the information verbally. The letter states a mailing label is enclosed to return the affected product. Customers are requested to amek the package as "recalled product" and include the XC or XH number. All products should be returned to the following shipping address: Bard Peripheral Vascular, Inc. 1415 W. 3rd St. Tempe, AZ 85281

Details

Recall number
Z-0786-2016
Initiated
January 5, 2016
Posted by FDA
February 9, 2016
Terminated
March 25, 2016
Product code
DSY – Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Quantity
4 units
Distribution
Distributed in the states of AL, TX, and CA.
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0785-2016 Class IIBard Distaflo Vascular Bypass Graft; Product Code: DF6006SC; 8 mm x 60 cm. Intended for…Size mislabeling; the graft is labeled as a 6 mm diameter graft when the actual packaged product is an 8 mm… 01/05/2016Terminated