Recall Z-0790-2022

Bard Peripheral Vascular Inc · initiated January 20, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Safe-T PLUS Pediatric/Infant Lumbar Puncture Tray, 22 g x 1.5" Spinal Needle with Manometer

Codes / lots: Catalog Number: 4302CSP Lot Numbers/UDI (GTIN, DI + PI): 0001412493 (01)10885403107788(17)220331(10)0001412493 0001415324 (01)10885403107788(17)220430(10)0001415324 0001416351 (01)10885403107788(17)220430(10)0001416351 0001417052 (01)10885403107788(17)220430(10)0001417052 0001417550 (01)10885403107788(17)220430(10)0001417550 0001421010 (01)10885403107788(17)220531(10)0001421010 0001422378 (01)10885403107788(17)220630(10)0001422378 0001425085 (01)10885403107788(17)220630(10)0001425085 0001426188 (01)10885403107788(17)220630(10)0001426188 0001426941 (01)10885403107788(17)220630(10)0001426941 0001428873 (01)10885403107788(17)220630(10)0001428873 0001431044 (01)10885403107788(17)220630(10)0001431044 0001433024 (01)10885403107788(17)220831(10)0001433024 0001433846 (01)10885403107788(17)220930(10)0001433846 0001437468 (01)10885403107788(17)221031(10)0001437468 0001438331 (01)10885403107788(17)221031(10)0001438331 0001439310 (01)10885403107788(17)221031(10)0001439310 0001440531 (01)10885403107788(17)221031(10)0001440531

Reason for recall

Due to lumbar puncture tray manometer failure mode of leaking and mating issues.

FDA-determined cause: Process control

Action taken

Between 01/28 and 02/01/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via email and FedEx 2-day mailings to customers informing them that between 07/02 to 12/15/2021 the firm received 59 customer complaints for lumbar puncture tray manometers with reported failure modes of leaking and mating issues. Customers are asked to take the following actions: 1. Check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facility s destruction process. 2. Share this notification with all users within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If Customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form to the Recalling Firm contact noted on the form confirming acknowledgement of the correction notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process their credit. 5. Indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory was destroyed. 6. Contact their Recalling Firm's representative to assist in this process or they may also contact the listed in the notification for further assistance. For further assistance contact: North American Regional Complaint Center - 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT

Details

Recall number
Z-0790-2022
Initiated
January 20, 2022
Product code
PWY – Lumbar Puncture Tray (Adult & Pediatric)
Quantity
20,400 devices
Distribution
U.S. Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S. (foreign) including states: Costa Rica and Hong Kong.
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0791-2022 Class IILUMBAR PUNCTURE TRAY PEDIATRIC 21G 2.5Due to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0792-2022 Class IILUMBAR PUNCTURE TRAY ADULT 20G X 3.5 DPDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0793-2022 Class IIADULT LUMBAR PUNCTURE DRUG FREE TRAY 20GDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0794-2022 Class IISAFE-T PLUS LUMBAR PUNCTURE TRAY ADULTDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0795-2022 Class IILUMBAR PUNCTURE TRAY ADULT 18G X 3.5Due to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0796-2022 Class IILUMBAR PUNCTURE TRAY ADULT 22G X3.5 STRLDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0797-2022 Class IISAFE-T PLUS LUMBARPUNCTURE TRAY ADULTDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0798-2022 Class IIATRAUMATIC ADULT LUMBAR PUNCTURE TRAYDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0799-2022 Class IILUMBAR PUNCTURE DRUG FREE TRAY 22GDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified
Z-0800-2022 Class IISAFE-T PLUS ATRAUMATIC ADULT LP TRAYDue to lumbar puncture tray manometer failure mode of leaking and mating issues. 01/20/2022Open, Classified