Recall Z-0793-2013
Hitachi Chemical Diagnostics Inc · initiated December 21, 2012
Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws
Product
OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France
Codes / lots: Catalog number: 85035. Lot number 42K283626, expiration May 31, 2013.
Reason for recall
There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.
FDA-determined cause: Process control
Action taken
Hitachi sent an Urgent Product Recall Notification letter dated December 21, 2012, to the affected customer. The letter identified the product, the problem, and the action to be taken by the customer. The customer was instructed to discontinue distribution and destroy any remaining inventory of the affected product at their location per their local waste ordinances, determine if any of their customers are using or have inventory of the affected product, rquest that their customers review all patient results reported with the use of the affectted product. The customer was instructed to tell their customers to discontinue use and destroy any remaining inventory per their local waste ordinances. The customer was instructed to retain a copy of the notification letter for their records. The customer was also instructed to send confirmation of the destruction within 5 days Customers were instructed to complete and return the attached Certificate of Destruction. For questions customers were instructed to contact their local area Customer Service. For questions regarding this recall 650-961-5501.
Details
- Recall number
- Z-0793-2013
- Initiated
- December 21, 2012
- Posted by FDA
- February 7, 2013
- Terminated
- May 16, 2013
- Product code
- DHB – System, Test, Radioallergosorbent (Rast) Immunological
- Quantity
- 20 kits
- Distribution
- Worldwide Distribution - France only
- Recalling firm
- Hitachi Chemical Diagnostics Inc, Mountain View, CA