Recall Z-0793-2013

Hitachi Chemical Diagnostics Inc · initiated December 21, 2012

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France

Codes / lots: Catalog number: 85035. Lot number 42K283626, expiration May 31, 2013.

Reason for recall

There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.

FDA-determined cause: Process control

Action taken

Hitachi sent an Urgent Product Recall Notification letter dated December 21, 2012, to the affected customer. The letter identified the product, the problem, and the action to be taken by the customer. The customer was instructed to discontinue distribution and destroy any remaining inventory of the affected product at their location per their local waste ordinances, determine if any of their customers are using or have inventory of the affected product, rquest that their customers review all patient results reported with the use of the affectted product. The customer was instructed to tell their customers to discontinue use and destroy any remaining inventory per their local waste ordinances. The customer was instructed to retain a copy of the notification letter for their records. The customer was also instructed to send confirmation of the destruction within 5 days Customers were instructed to complete and return the attached Certificate of Destruction. For questions customers were instructed to contact their local area Customer Service. For questions regarding this recall 650-961-5501.

Details

Recall number
Z-0793-2013
Initiated
December 21, 2012
Posted by FDA
February 7, 2013
Terminated
May 16, 2013
Product code
DHB – System, Test, Radioallergosorbent (Rast) Immunological
Quantity
20 kits
Distribution
Worldwide Distribution - France only
Recalling firm
Hitachi Chemical Diagnostics Inc, Mountain View, CA

Official FDA recall record