Recall Z-0801-2021

Ortho-Clinical Diagnostics, Inc. · initiated November 20, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.

Codes / lots: enGen Laboratory Automation Systems and VITROS Automation Solutions Using All TCAutomationTM configured with Thermo Scientific Aliquoter & Labeller Module and using Software Versions 4.6 and below

Reason for recall

A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.

FDA-determined cause: Software design

Action taken

On November 20, 2020, the firm sent Urgent Product Correction Notification letters to affected customers to inform them of the issue. Customers were reminded to follow product labelling regarding removing all tubes before performing maintenance or initializing the Aliquoter & Labeller Module. There should never be sample tubes present when initializing the Aliquoter & Labeller module. The firm's release of TCA Software version 4.7 is the planned mitigation for this issue. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.

Details

Recall number
Z-0801-2021
Initiated
November 20, 2020
Posted by FDA
January 7, 2021
Terminated
October 24, 2023
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
40 units
Distribution
Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.
Recalling firm
Ortho-Clinical Diagnostics, Inc., Rochester, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0800-2021 Class IIenGen Laboratory Automation System, enGen Track System, Catalog Number ENGEN - Product…A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to… 11/20/2020Terminated