Recall Z-0810-2022

Biomet, Inc. · initiated February 14, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BIOMET Trauma, [Item # 13182-22-26, DVR CROSSLOCK EXTRA LONG PLATE LEFT], [Item # 1318-12-126, DVR CROSSLOCK EXTRA LONG PLATE RIGHT], [Item # 1318-12-176, DVR CROSSLOCK EXTRA EXTRA LONG PLATE RIGHT]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Codes / lots: 6 lots: Lot #, UDI - [101220, (01)00887868446191(17)310623(10)101220], [101340, (01)00887868446207(17)310625(10)101340], [101450, (01)00887868446313(17)310630(10)101450], [101560, (01)00887868446191(17)310624(10)101560], [119180, (01)00887868446313(17)310630(10)119180], [154060, (01)00887868446313(17)310630(10)154060]

Reason for recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

FDA-determined cause: Under Investigation by firm

Action taken

A recall notification was sent via courier 24 Feb 2022 to distributors, surgeons, and hospital risk managers. Customers advised to locate and quarantine product, return to Biomet and identify any further customers receiving product. Recalled product is identified using specific lot and item numbers. Additionally, surgeons and hospital risk managers are to follow any specified patient monitoring instructions and return the Certificate of Acknowledgement to Zimmer Biomet. If you have further questions or concerns, please call customer service at 574-371-3071 or send questions via email to CorporateQuality.PostMarket@zimmerbiomet.com

Details

Recall number
Z-0810-2022
Initiated
February 14, 2022
Terminated
September 30, 2024
Product code
JDR – Staple, Fixation, Bone
Quantity
38
Distribution
Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands
Recalling firm
Biomet, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0809-2022 Class IIBIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item…Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile… 02/14/2022Terminated
Z-0811-2022 Class IIBIOMET TRAUMA A.L.P.S. FIBULA COMPOSITE LOCKING PLATE 8 HOLE, Item # 856204008. Devices…Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile… 02/14/2022Terminated
Z-0812-2022 Class IIBIOMET TRAUMA Drill Bit, Tap, Guide Pin: [DVR Drill Bit, Item # 212000022, DVR 2.2MM…Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile… 02/14/2022Terminated