Recall Z-0816-2014

Radiometer America Inc · initiated October 15, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ABL90 FLEX Analyzer Radiometer Medical ApS, kandevej 21 Brnshj, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.

Codes / lots: Lot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010

Reason for recall

RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

FDA-determined cause: Other

Action taken

Radiometer sent an Urgent Medical Device Recall letter dated November 5, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check all of their ABL90 FLEX analyzers and ensure that the analyzer is placed on a horizontal platform/table. If the analyzer is placed on a roll stand, please ensure that the support plate is aligned horizontally and locked in this position. Customers were asked to complete and fax the accompanying acknowledgment form in order to confirm their receipt of this field safety notice. Customers with questions were instructed to contact Radiometer America Technical Support at 1-800-736-0600, option 4. For questions regarding this recall call 440-871-8900.

Details

Recall number
Z-0816-2014
Initiated
October 15, 2013
Posted by FDA
January 23, 2014
Terminated
June 19, 2014
Product code
JGS – Electrode, Ion Specific, Sodium
Quantity
Total 3414 (US 524, CA 31, ROW 2859)
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South…
Recalling firm
Radiometer America Inc, Westlake, OH