Recall Z-0817-2024
American Contract Systems, Inc. · initiated November 29, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL
Codes / lots: a) JRHC45G, UDI/DI 00191072164240, Lot Numbers: 912231, exp. 3/29/2024; 920231, exp. 3/21/2024; 952231, exp. 2/17/2024; 963231, exp. 2/6/2024; 976231, exp. 6/2/2023; 654221, exp. 12/12/2023; 714221, exp. 10/13/2023; 717221, exp. 6/7/2023; 766221, exp. 8/22/2023; 814221, exp. 7/5/2023; b) LLTV10, UDI/DI 00191072183255, Lot Numbers: 644221, exp. 12/22/2023; 973231, exp. 1/27/2024; 952231, exp. 2/17/2024; 924231, exp. 3/17/2024; c) SJCA21F, UDI/DI None, Lot Numbers: 666221, exp. 11/30/2023; 701221, exp. 10/26/2023; 747221, exp. 9/10/2023; 779221, exp. 8/9/2023; 807221, exp. 7/12/2023; 846221, exp. 6/3/2023; 919231, exp. 3/22/2024; 954231, exp. 2/15/2024; 988231, exp. 1/12/2024; d) SJCH22F, UDI/DI None, Lot Numbers: 670221, exp. 11/26/2023; 708221, exp. 10/19/2023; 747221, exp. 9/10/2023; 795221, exp. 7/24/2023; 834221, exp. 6/15/2023; 873221, exp. 5/7/2023; 937231, exp. 3/4/2024; 991231, exp. 1/9/2024; e) UDIR58AL, UDI/DI None, Lot Numbers: 682221, exp. 11/14/2023; 638221, exp. 7/31/2023; 963231, exp. 2/6/2024; 925231, exp. 3/16/2024; 883231, exp. 10/30/2023
Reason for recall
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
FDA-determined cause: Process control
Action taken
Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
Details
- Recall number
- Z-0817-2024
- Initiated
- November 29, 2023
- Posted by FDA
- January 26, 2024
- Product code
- OES – Cardiac Catheterization Kit
- Quantity
- 4394 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- American Contract Systems, Inc., Temple Terrace, FL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0816-2024 Class II | Custom procedural convenience kits and trays, anesthesia, labeled as: a) HEART…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0818-2024 Class II | Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0819-2024 Class II | Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0820-2024 Class II | Custom procedural convenience kits and trays, ear, nose & throat, labeled as: a) T AND A…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0821-2024 Class II | Custom procedural convenience kits and trays, gastroenterological & urological, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0822-2024 Class II | Custom procedural convenience kits and trays, gastroenterological & urological, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0823-2024 Class II | Custom procedural convenience kits and trays, urological, labeled as: a) UROLOGY DAVINCI…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0824-2024 Class II | Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a)…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0825-2024 Class II | Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a)…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0826-2024 Class II | Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0827-2024 Class II | Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0828-2024 Class II | Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0829-2024 Class II | Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0830-2024 Class II | Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0831-2024 Class II | Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0832-2024 Class II | Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |
| Z-0833-2024 Class II | Custom procedural convenience kits and trays, Othopedic use, labeled as: a) HIP…During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and… | 11/29/2023Open, Classified |