Recall Z-0818-2018

Zimmer Biomet, Inc. · initiated August 22, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Ziptight Ankle Syndesmosis Fixation Device, Titanium Implant Titanium Implant for surgical repairs of soft tissue.

Codes / lots: Item No. 905759, Lot No. (UDI No.): 58860 ((01)00880304476103(17)220523(10)058860), 58880 ((01)00880304476103(17)220608(10)058880), 58890 ((01)00880304476103(17)220608(10)058890), 58910 ((01)00880304476103(17)220608(10)058910), 116030 ((01)00880304476103(17)220519(10)116030), 169780 ((01)00880304476103(17)220609(10)169780), 169790 ((01)00880304476103(17)220608(10)169790), 213450 ((01)00880304476103(17)220609(10)213450), 213480 ((01)00880304476103(17)220607(10)213480), 374420 ((01)00880304476103(17)220319(10)374420), 392530 ((01)00880304476103(17)220610(10)392530), 476670 ((01)00880304476103(17)220607(10)476670), 503830 ((01)00880304476103(17)220608(10)503830), 614910 ((01)00880304476103(17)220421(10)614910), 670640((01)00880304476103(17)220621(10)670640), 728190 ((01)00880304476103(17)220606(10)728190), 984070 ((01)00880304476103(17)220512(10)984070), 984080 ((01)00880304476103(17)220607(10)984080)

Reason for recall

Ziptight Ankle Syndesmosis Fixation was packaged without the tip protectors.

FDA-determined cause: Packaging process control

Action taken

Customers were notified via letter on about 08/22/2017. Instructions for distributors, sales representatives, and distributor operation managers include to ensure affected personnel are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return affected product, and identify any additional hospitals or surgeons that should be notified. Instructions for surgeons and risk managers include to ensure affected personnel are aware of the recall, identify and quarantine any affected product on hand, complete and return the Certificate of Acknowledgement, return any affected product. Questions or concerns can be addressed to the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST, Monday through Friday. Additionally, questions may be sent by email to corporatequality.postmarket@zimmerbiomet.com.

Details

Recall number
Z-0818-2018
Initiated
August 22, 2017
Terminated
May 12, 2020
Product code
MBI – Fastener, Fixation, Nondegradable, Soft Tissue
Quantity
885 units total
Distribution
Distributed domestically to AK, AL, AR, AZ, CA, CO, FL, ID, IN, KY, MA, MD, ME, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI. Distributed internationally to ARGENTINA, AUSTRALIA, CANADA, COSTA RICA, JAPAN, NETHERLANDS.
Recalling firm
Zimmer Biomet, Inc., Warsaw, IN

Official FDA recall record