Recall Z-0831-2012

Gulf Medical Fiberoptics, Inc. · initiated May 13, 2011

Terminated

Product

Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, and GMF-H106. The Surgical Headlight is a passive luminaire. It is illuminated by a source of light by way of the fiber optics and provides supplemental lighting for medical and surgical procedures.

Codes / lots: Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, GMF-H106.

Reason for recall

Gulf Medical Fiberoptics, Inc. recalled their Surgical Headlight because the cleaning instructions for the exterior surfaces of the headlight , headgear and fiber cable with 70% isopropyl alcohol had not been validated.

FDA-determined cause: Process control

Action taken

On May 16, 2011, Gulf Medical Fiberoptics, Inc. issued to all customers an updated letter to their May 13, 2011 "Field Correction of Instructions For Use" informing them to discontinue the use of 70% isopropyl to clean the surfaces of the device. Enclosed with the letter was an updated Surgical Headlight User Guide with instructions to discard/shred/recycle the prior guide. The firm asked customers to return a form acknowledging the receipt and action from the letter. Customers with questions should contact Customer Service department manager Jacqueline Armand at 813-855-6618 ext. 427.

Details

Recall number
Z-0831-2012
Initiated
May 13, 2011
Posted by FDA
January 20, 2012
Terminated
August 28, 2012
Product code
FST – Light, Surgical, Fiberoptic
Quantity
381 units
Distribution
Worldwide Distribution - USA, including the states of FL, WI, WA, AL, CA, MO, MI, OH, VA, CO, TX, TN, NC and the countries of Turkey, Brazil, Belgium, South Korea, Panama, Mexico, Ireland, Taiwan (ROC), Italy, Argentina, India, Spain, Germany, and Venezuela.
Recalling firm
Gulf Medical Fiberoptics, Inc., Oldsmar, FL

Official FDA recall record