Recall Z-0831-2019

AtriCure, Inc. · initiated January 22, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

COBRA FUSION 150 Ablation System, Catalog # (REF) 700-001S

Codes / lots: UDI - 00818354012828 Lot Numbers - All lots within expiry

Reason for recall

There is a potential for patients undergoing ablation to experience an elevated risk for TE; during stand-alone off-pump procedures where cardio pulmonary bypass (CPB) was not utilized.

FDA-determined cause: Unknown/Undetermined by firm

Action taken

On January 22, 2019, AtriCure distributed URGENT Advisory Notices to customers via certified mail. AtriCure is providing modifications to the existing Cobra Fusion Instruction for Use (IFU), to address contributing TE Factors associated with stand-alone off-pump procedures. These are in addition to the warnings and precautions that are present within the current IFU. Modifications to Warnings: When utilizing the Cobra Fusion device in a stand-alone off-pump (without CPB) procedure, the following should be considered: - Physicians should consider a comprehensive anti-coagulation protocol including pre-operative, intra-operative and post-operative anti-coagulation management to prevent potential thromboemboli. Modifications to Precautions: - Post-operative anti-coagulation therapy for protection against thromboemboli is inclusive of the bridging period between the end of the procedure and until effective therapeutic levels of Oral Anticoagulation (OAC) are achieved. Customers Should Take the following Action: - Read and follow the revised Instructions for Use IFU enclosed with the notification letter when using the Cobra Fusion. - Report any post ablation thromboembolic event with the Cobra Fusion, to AtriCure by phone at 1-866-349-2342 (select option 6) or e-mail to pcomplaints@atricure.com. - Return the completed Acknowledgement Form. Contact Information: If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-644-4245) from 9-6pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for review promptly. **Press Release will be issue depended on CDRH comment/review**

Details

Recall number
Z-0831-2019
Initiated
January 22, 2019
Terminated
May 12, 2020
Product code
OCL – Surgical Device, For Cutting, Coagulation, And/Or Ablation Of Tissue, Including Cardiac Tissue
Quantity
175 units
Distribution
AL, AR, AZ, CA, CT, CO, FL, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA & WI Internationally - Germany, Italy, Netherlands, France, Austria, United Kingdom, Poland, Czech Republic, United Arab Emirates, Sweden, Switzerland, Kuwait, Belgium and Latvia
Recalling firm
AtriCure, Inc., Mason, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0829-2019 Class IICOBRA FUSION 50, Ablation System, Catalog # (REF) 700-002There is a potential for patients undergoing ablation to experience an elevated risk for TE; during… 01/22/2019Terminated
Z-0830-2019 Class IICOBRA FUSION 150 Ablation System, Catalog # (REF) 700-001There is a potential for patients undergoing ablation to experience an elevated risk for TE; during… 01/22/2019Terminated
Z-0832-2019 Class IICOBRA FUSION 150 Ablation System (International Only), Catalog # (REF) 700-001MIThere is a potential for patients undergoing ablation to experience an elevated risk for TE; during… 01/22/2019Terminated