Recall Z-0839-2011

AngioScore Inc. · initiated November 16, 2010

Terminated

Product

AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm, 5.0 mm x 20 mm balloon size, Part (REF) number: 2076-5020.

Codes / lots: Part (REF) number: 2076-5020: All lot codes

Reason for recall

Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.

FDA-determined cause: Device Design

Action taken

Medical Device Recall notifications were sent to all domestic consignees on Nov 16, 2010 by overnight mail. Foreign distributors, other than Japan, were sent recall letters by e-mail or international overnight delivery on November 16, 2010. The Japanese distributor received notification and notified their customers on November 10, 2010.

Details

Recall number
Z-0839-2011
Initiated
November 16, 2010
Posted by FDA
January 4, 2011
Terminated
October 4, 2011
Product code
DQY – Catheter, Percutaneous
Quantity
17682 units, all sizes and lots
Distribution
Worldwide Distribution: USA, China, Finland, Italy, Japan, Norway, Saudi Arabia, Spain, Sweden, Switzerland, UAE, United Kingdom
Recalling firm
AngioScore Inc., Fremont, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0838-2011 AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm…Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end… 11/16/2010Terminated
Z-0840-2011 AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm…Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end… 11/16/2010Terminated
Z-0841-2011 AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 90 cm…Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end… 11/16/2010Terminated
Z-0842-2011 AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 50 cm…Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end… 11/16/2010Terminated