Recall Z-0839-2014

DeRoyal Industries Inc · initiated September 17, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Codes / lots: Lot Numbers: 32901129, 32980912, 33015541, 33024746, 33105239

Reason for recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

FDA-determined cause: Device Design

Action taken

DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.

Details

Recall number
Z-0839-2014
Initiated
September 17, 2013
Posted by FDA
January 27, 2014
Terminated
June 5, 2017
Product code
ILO – Pack, Hot Or Cold, Water Circulating
Quantity
25 units
Distribution
Nationwide Distribution.
Recalling firm
DeRoyal Industries Inc, New Tazewell, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0834-2014 Class IIAquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal…The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the… 09/17/2013Terminated
Z-0835-2014 Class IIDeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF…The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the… 09/17/2013Terminated
Z-0836-2014 Class IIDeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS…The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the… 09/17/2013Terminated
Z-0837-2014 Class IIDeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF…The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the… 09/17/2013Terminated
Z-0838-2014 Class IIDeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF…The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the… 09/17/2013Terminated