Recall Z-0839-2019

Randox Laboratories Ltd. · initiated December 4, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Codes / lots: All lot codes

Reason for recall

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

FDA-determined cause: Device Design

Action taken

Randox the US distributor contacted customers by email on 12/4/18 to advise of problem and provided an update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments.The instrument testing order should be reviewed in line with the updated technical bulletin. A revised letter issued 12 Feb 2019 to advise "reagent would be observed as inconsistencies in Quality control recovery which may lead to a delay in running patient samples or erroneous elevated / depressed test results".

Details

Recall number
Z-0839-2019
Initiated
December 4, 2018
Terminated
June 18, 2020
Product code
CJY – Azo Dye, Calcium
Quantity
225 kits
Distribution
WV, PR
Recalling firm
Randox Laboratories Ltd., Crumlin, North

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0840-2019 Class IIRandox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo…Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover… 12/04/2018Terminated