Recall Z-0842-2025

Augustine Temperature Management, LLC · initiated March 14, 2022

Class II Completed

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V; 2) WC71V-DEMO; 3) WC71V-R; used in conjunction with Warming Blankets and Mattresses, veterinary use.

Codes / lots: 1) UDI/DI 00850011479247, Serial Numbers: 20090015, 20090016, 20090017, 20090018, 20090020, 20090021, 20090022, 20090023, 20090024, 20090025, 20090026, 20090027, 20090028, 20090029, 20090030, 20090031, 20090032, 20090033, 20090034, 20090035, 20090036, 20090037, 20090038, 20090039, 20090040, 20090041, 20090042, 20090043, 20090044, 20100030, 20100031, 20100032, 20100033, 20100034, 20100035, 20100036, 20100037, 20100038, 20100040, 20100041, 20100042, 20100043, 20100044, 20100045, 20100046, 20100047, 20100048, 20100049, 20100050, 20100051, 20100052, 20100053, 20100054, 20100055, 20100056, 20100057, 20100058, 20100059, 20100091, 20100092, 20100093, 20100094, 20100095, 20100096, 20100097, 20100098, 20100099, 20100100, 20100101, 20100103, 20110000, 20110001, 20110002, 20110003, 20110004, 20110005, 20110006, 20110006, 20110007, 20110008, 20110009, 20110010, 20110011, 20110012, 20110013, 20110014, 20110015, 20110016, 20110017, 20110018, 20110019, 20110020, 20110021, 20110022, 20110023, 20110024, 20110025, 20110026, 20110027, 20110028, 20110029, 20110030, 20110031, 20110032, 20110033, 20110034, 20110035, 20110036, 20110037, 20110038, 20110039, 20110040, 20110041, 20110042, 20110043…

Reason for recall

HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".

FDA-determined cause: Device Design

Action taken

Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.

Details

Recall number
Z-0842-2025
Initiated
March 14, 2022
Posted by FDA
January 10, 2025
Product code
DWJ – System, Thermal Regulating
Quantity
2271 units
Distribution
Worldwide
Recalling firm
Augustine Temperature Management, LLC, Minnetonka, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0839-2025 Class IIAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller…HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model… 03/14/2022Completed
Z-0840-2025 Class IIAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller…HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model… 03/14/2022Completed
Z-0841-2025 Class IIHillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with…HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model… 03/14/2022Completed
Z-0843-2025 Class IIAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller…HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model… 03/14/2022Completed