Recall Z-0844-2022

Microbiologics Inc · initiated March 3, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Pneumonia (33 Targets) Control Panel, unassayed quality control material, containing: Positive Control 1 - Semi-Quantitative Bacteria Positive Control 2 - Viruses and Atypical Bacteria

Codes / lots: Catalog Number 8244, UDI: 70845357043664, Lot Number (Expiration Date) (New Expiration Date): 8244 01 (07/31/2022) (07/26/2021), 8244 02 (08/31/2022) (08/02/2021)

Reason for recall

Shelf-life for the 8244 Pneumonia (33 Targets) Control Panel determined to be 6 months instead of 18 months.

FDA-determined cause: Incorrect or no expiration date

Action taken

Microbiologics notified customers on 03/03/2022 via letter titled "Urgent Medical Device Recall Re: 8244 Pneumonia (33 Targets) Control Panel". Instructions included to review lab procedures to understand how the information affects usage, discard expired product, complete the response form provided, return the response form to recall@microbiologics.com, and keep the letter for consignees' records. Questions or concerns can be directed to the Recall Support team at 320.229.7073 or recall@microbiologics.com.

Details

Recall number
Z-0844-2022
Initiated
March 3, 2022
Terminated
January 16, 2024
Product code
OHQ – Multi-Analyte Controls Unassayed
Quantity
13 units
Distribution
US distribution to Alabama, Georgia, Illinois, Montana, Pennsylvania, Tennessee, Texas, and Wisconsin
Recalling firm
Microbiologics Inc, Saint Cloud, MN