Recall Z-0849-2019
Baxter Healthcare Corporation · initiated December 11, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
AMIA Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing Peritoneal dialysis.
Codes / lots: Product code 5C9320. All serial numbers are affected for devices distributed from 02/11/2016 through 11/27/2018.
Reason for recall
Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
FDA-determined cause: Software design
Action taken
Baxter Healthcare Corporation notified providers on about 12/11/2018 of the recall concerning the AMIA Automated Peritoneal Dialysis System, via "Urgent Medical Device Correction" letter. Patients were notified on about 1/14/2018 via "Urgent Medical Device Correction" letter. Providers were instructed to continue using the AMIA Automated PD System however, for patients who are prescribed a Cycle-Based therapy, to change these therapies to Time-Based by following the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. In Time-Based therapy, to maintain a fixed dwell time, the Smart Dwell treatment option needs to be Disabled. To disable the Smart Dwell treatment option, follow the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. Questions can be directed to Baxter's Renal Clinical Helpline at 888-736-2543, option 2, Monday through Friday, 8:00 am to 4:30 pm (CT). Providers were informed that Baxter would contact them when the software update is available and provide further instructions on how to download the upgrade remotely. They were also instructed to complete and return the Baxter Customer Reply Form and to inform customers if the device was further distributed to other facilities or departments within their institution. Patients were instructed to continue using the AMIA Automated PD System however, they were directed to check the Programming Method to ensure they have not been prescribed Cycle-Based therapy. The programming method can be verified by following section 10.3 of the Patient Guide: 1. touch "GO" on the home screen and the Confirm Treatment screen should appear, 2. touch "Program Settings" to view the program settings details and the Program Details screen will appear, 3. display the "Program Method" by navigating to Page 3 by pressing the down arrow at the bottom of the screen. If the Programming Method is Time-Based or Tidal, the iss
Details
- Recall number
- Z-0849-2019
- Initiated
- December 11, 2018
- Terminated
- February 12, 2021
- Product code
- FKX – System, Peritoneal, Automatic Delivery
- Quantity
- 7,688 devices
- Distribution
- Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Canada and Japan.
- Recalling firm
- Baxter Healthcare Corporation, Deerfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0850-2019 Class II | Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic…Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a… | 12/11/2018Terminated |