Recall Z-0852-2016

Stryker Spine · initiated December 16, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.

Codes / lots: Catalog number 900301, Lot #01131543, 01261508, 10151404, 10271408, 12161403, 12161404, 09021404, 01261509, 10271407, 01261507, 10071415, 05011403, 06241407, 08071402, 10091305, 03041403, 01271412, 07281401, and 11061303. Catalog number 905103, 01131542, 11241408, 12161405, 12161406, 0211501, 01211502, 11241406, 02031403, 06241408, 07161407, 09101402, 09281401, 10251303, 05301402, 11241407, and 12101305.

Reason for recall

There were two potential interference conditions identified with the way the tubing set attaches to the inserter.

FDA-determined cause: Device Design

Action taken

Stryker Spine sent an Urgent Product Removal letter dated December 17, 2015 to affected customers via Fed Ex priority overnight. The letter identified the affected product, problem and actions to be taken. Customers are asked to follow instructions included in the letter. The customer response form should be completed and sent back to email via Spine-RegulatoryActions@Stryker.com or Fax 855-632-9049. For questions call (201)749-8389.

Details

Recall number
Z-0852-2016
Initiated
December 16, 2015
Terminated
August 1, 2016
Product code
MAX – Intervertebral Fusion Device With Bone Graft, Lumbar
Quantity
413 units
Distribution
US Nationwide Distribution
Recalling firm
Stryker Spine, Allendale, NJ

Official FDA recall record