Recall Z-0853-2007

Diagnostica Stago, Inc. · initiated March 13, 2007

Terminated

Product

STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.

Codes / lots: Lot -- Exp: 052783 -- April 2007; 061242 -- November 2007; 061981 -- February 2008.

Reason for recall

A dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.

FDA-determined cause: Other

Action taken

Customers were notified on March 13, 2007 by letter, first class mail.

Details

Recall number
Z-0853-2007
Initiated
March 13, 2007
Posted by FDA
July 3, 2007
Terminated
January 16, 2009
Product code
GGP – Test, Qualitative And Quantitative Factor Deficiency
Quantity
1,098 units
Distribution
Hospitals and laboratories Nationwide
Recalling firm
Diagnostica Stago, Inc., Parsippany, NJ

Official FDA recall record