Recall Z-0853-2015

Zimmer, Inc. · initiated December 4, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.

Codes / lots: Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754

Reason for recall

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

FDA-determined cause: Process design

Action taken

Notifications will be sent to the affected distributors and hospital risk managers with instructions for returning the affected product the week of December 8, 2014. Recalled product will be quarantined and returned to Zimmer Sales Representatives to be destroyed. The public contact for the recall is: Consumer Relations Call Center Zimmer, Inc. P O Box 708 Warsaw, IN 46581-0708 consumer@zimmer.com Phone: 1-800-447-5633 Fax: 574-371-8755

Details

Recall number
Z-0853-2015
Initiated
December 4, 2014
Posted by FDA
December 23, 2014
Terminated
May 13, 2015
Product code
JWH – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Quantity
83 units
Distribution
Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Recalling firm
Zimmer, Inc., Warsaw, IN