Recall Z-0860-2019

Philips North America, LLC · initiated January 9, 2019

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

PageWriter TC50 Government Bundle, Product 860429, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Codes / lots: USN1800002 USN1800003 USN1800004

Reason for recall

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

FDA-determined cause: Use error

Action taken

On December 8, 2018, Philips issued an Urgent Medical Device Correction letter to its customers, informing them of the product issue. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction/Field Safety Notice This includes requesting that users: Carefully read the PageWriter TC Service Manual Addendum and to review this information with all staff members who are responsible for device management of the Philips PageWriter TC cardiographs Promptly determine the number of cycles and the State of Health (SOH) on each of the affected Philips PageWriter TC cardiographs (TC20/30/50/70), as specified in the Service Manual Addendum. Replace the battery if the number of cycles is greater than 300 and/or if the SOH is less than 80. Customers are instructed to complete and return the response card provided with the notice. Philips is instructing customers that once it is determined the battery is not in need of replacement, or once the battery is replaced, the PageWriter PC Cardiograph is safe to continuing using. Philips plans to release a customer installable, software update for PageWriter TC cardiograph (TC20/30/50/70) that will provide alerts to assist users in managing the battery replacement cycle. Users will be notified when the customer installable software update is available. The software will provide notification the customer when it is time to replace the battery, based on charge-discharge cycle count and the batteries state of health (SOH).

Details

Recall number
Z-0860-2019
Initiated
January 9, 2019
Product code
DPS – Electrocardiograph
Quantity
3
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recalling firm
Philips North America, LLC, Andover, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0853-2019 Class IIPageWriter TC70, Product 860315, Software revisions up to and including A.07.05.22 used…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0854-2019 Class IIPageWriter TC50, Product 860310, Software revisions up to and including A.07.05.22 used…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0855-2019 Class IIPageWriter TC30, Product 860306, Software revisions up to and including A.07.05.22 to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0856-2019 Class IIPageWriter TC20, Product 860332, Software revisions up to and including A.07.05.22 to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0857-2019 Class IIPageWriter TC70 w/trolley Government Bundle, Product 860352, Software revisions up to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0858-2019 Class IIPageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0859-2019 Class IIPageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified
Z-0861-2019 Class IIPageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up to…Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to… 01/09/2019Open, Classified