Recall Z-0864-2016

EKOS Corporation · initiated January 21, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.

Codes / lots: Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010

Reason for recall

The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.

FDA-determined cause: Error in labeling

Action taken

All affected consignees were contacted by telephone January 20- 22, 2016. The URGENT VOLUNTARY PRODUCT RECALL letter, dated January 21, 2016, was sent to consignees. Consignees were advised that replacement devices were shipped to their facility on January 21, 2016 and should arrive there on January 22, 2016. Consignees with questions can call: Jocelyn Kersten Vice President, Regulatory and Clinical Affairs EKOS Corporation, a BTG International Group Company 425-415-3132 Jocelyn.kersten@ekoscorp.com

Details

Recall number
Z-0864-2016
Initiated
January 21, 2016
Posted by FDA
February 24, 2016
Terminated
March 15, 2016
Product code
KRA – Catheter, Continuous Flush
Quantity
10 units
Distribution
Distributed in the states of AZ, CA, MA, MI, OH, and VA.
Recalling firm
EKOS Corporation, Bothell, WA

Official FDA recall record