Recall Z-0885-2016

Biomet, Inc. · initiated February 16, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

HipLOC CHS Plate/Lag Screw Introducer

Codes / lots: Catalog Number: 35-000263; and Lot Number Identification: 1807188, 1807189, 1943563, and 2577847.

Reason for recall

A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.

FDA-determined cause: Process change control

Action taken

Biomet sent an " URGENT MEDICAL DEVICE RECALL NOTICE" dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation. Remove the affected product from circulation. Review this notification and ensure that all affected personnel are aware of its contents. Carefully follow the instructions on the enclosed Response Form. Email a copy of the response form to CPWARFieldAction@zimmerbiomet.com prior to return of product. You do not need to complete the Biomet UK response form. Use priority carrier for your shipment. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: "Online: www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or "Call (800)FDA-1088 Thank you in advance for your assistance and prompt attention. On behalf of Zimmer Biomet, I apologize for any inconvenience this may cause. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.

Details

Recall number
Z-0885-2016
Initiated
February 16, 2016
Posted by FDA
February 25, 2016
Terminated
February 14, 2017
Product code
LXH – Orthopedic Manual Surgical Instrument
Quantity
23 units
Distribution
US Distribution to the states of : MO, UT, AZ, IL, PA, KY, MI, NC, NJ, MN and GA., and Internationally to The Netherlands.
Recalling firm
Biomet, Inc., Warsaw, IN

Official FDA recall record