Recall Z-0886-2016

Teleflex Medical · initiated February 2, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Auto Endo5 Hem-o-lok Automatic Clip Applier, Medium Large Polymer Clips, Rx Only, Sterile, REF 543965 Product Usage: The Weck Auto Endo5 hem-o-lok ML automatic endoscopic ligating clip appliers are indicated for use as delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips. These appliers are designed for use with 5/5.5mm cannulas. Hem-o-lok Ligating Clips are intended for use in procedures involving ligation of vessels or tissue structures.

Codes / lots: 73J1500065, 73J1500066, 73J1500280, 73J1500440, 73J1500565, 73J1500566, 73K1500193

Reason for recall

The clip in the applier may be missing a boss. A boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.

FDA-determined cause: Nonconforming Material/Component

Action taken

Teleflex sent an Urgent Medical Device Recall Notification letter dated February 2, 2016 to customers. The letter identified the affected product, problem and actions to be taken. A separate letter was sent to distributors advising them to conduct a sub-recall. Customers were instructed to: 1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog numbers listed in the letter. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1- 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com.This will document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will document receipt of the letter. For questions contact your local sales representative or Customer Service at 1-866-246-6990

Details

Recall number
Z-0886-2016
Initiated
February 2, 2016
Posted by FDA
February 25, 2016
Terminated
August 11, 2017
Product code
FZP – Clip, Implantable
Quantity
543 ea.
Distribution
Worldwide Distribution in the state of CA, MI, NC, PA, TX, VA, and the countries of Canada, Korea & Belgium
Recalling firm
Teleflex Medical, Research Triangle Park, NC

Official FDA recall record