Recall Z-0887-2022
Maquet Cardiovascular, LLC · initiated February 9, 2022
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
Codes / lots: UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.
Reason for recall
Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
FDA-determined cause: Mixed-up of materials/components
Action taken
On February 14, 2022, a customer notification letter was issued via FedEx. Customers are instructed to quarantine and return all affected product and notify all users within the hospital/facility. Please note affected lot numbers appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number, which is etched onto all 3 device components on side of handle and on underside above each pin. Customers are to complete and return the response form by email or fax. Distributors are to forward this document to customers for appropriate action. If you have any questions, please contact your Maquet Cardiovascular, LLC /Getinge representative or call the Getinge Customer Service at (888) 880-2874, Monday through Friday, between the hours of 6:00a.m. and 5:00 p.m. (Pacific Standard Time)
Details
- Recall number
- Z-0887-2022
- Initiated
- February 9, 2022
- Product code
- DWS – Instruments, Surgical, Cardiovascular
- Quantity
- 60
- Distribution
- Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.
- Recalling firm
- Maquet Cardiovascular, LLC, Wayne, NJ