Recall Z-0896-2013

Orthofix, Inc · initiated February 1, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Codes / lots: Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers

Reason for recall

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

FDA-determined cause: Device Design

Action taken

Orthofix sent an Urgent Medical Device Recall Notification dated February 1, 2013 to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to identify and remove the affected products from inventory immediately and follow the actions listed in the enclosed Product Return Instructions. For questions or concerns contact Regulatory Affairs at 214-937-2100.

Details

Recall number
Z-0896-2013
Initiated
February 1, 2013
Posted by FDA
February 28, 2013
Terminated
July 1, 2013
Product code
KWP – Appliance, Fixation, Spinal Interlaminal
Quantity
565 total (287 PN 52-6800 and 278 PN 52- 6700)
Distribution
Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.
Recalling firm
Orthofix, Inc, Lewisville, TX