Recall Z-0898-2017

Teleflex Medical · initiated October 27, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Mucosal Atomization Device (MAD), Intranasal Mucosal Atomization Device with 3mL Syringe and Vial Adapter; Product Code MAD100, MAD100OS MAD110, MAD110OS, MAD130, MAD130OS, MAD140, MAD140OS, MAD300, MAD300B. Manufactured By: Wolfe-Tory Medical, Inc., Teleflex Medical. For atomization of topical solutions across the naso and oropharyngeal mucosal membranes

Codes / lots: Product code MAD100, Batch/Lot numbers 160105, 160137, 160302, 160321, 160402, 160435, 160506, 160523, 160609, 160620, 160707, 160802, 160813; Product code MAD100OS, Batch/Lot numbers 160322, 160524, 160630; Product code MAD110, Batch/Lot numbers 160217, 160507; Product code MAD110OS, Batch/Lot numbers 160240, 160312; Product code MAD130, Batch/Lot numbers 160107, 160138, 160517; Product code MAD130OS, Batch/Lot numbers 160436, 160803; Product code MAD140, Batch/Lot numbers 160125, 160218, 160437, 160610, 160801; Product code MAD140OS, Batch/Lot numbers 160226, 160438, 160727; Product code MAD300, Batch/Lot numbers 160108, 160117, 160126, 160145, 160146, 160200, 160219, 160225, 160231, 160300, 160313, 160327, 160400, 160409, 160422, 160432, 160440, 160500, 160518, 160602, 160611, 160621, 160631, 160701, 160708, 160718, 160728, 160800, 160804, 160814, 160816, 160823; Product code MAD300B, Batch/Lot number 160410.

Reason for recall

Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.

FDA-determined cause: Process control

Action taken

Teleflex sent an Urgent Medical Device Recall Notification letter on/about October 27, 2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and quarantine affected products. Customers were asked to complete the enclosed Recall Acknowledgement Form and fax it to Teleflex Customer Service or email to recalls@teleflex.com. For questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990. Teleflex sent an Amended Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to continue to return affected product for a full refund per the procedure set forth in the letter of October 27, 2016. Some customers have indicated that, due to medical necessity, they plan to continue using the affected lots rather than return them. Teleflex is advising all such customers to follow the supplemental instructions for Use included as Attachment B. These supplemental instructions allow non-destructive testing of each unit prior to the procedure to determine if it is defective. Institutions that intend to continue using affected lots with this supplemental testing, please send the acknowledgment form to recall@teleflex.com with their name and contact information and the lots that they will continue to use, or call 866-246-6990. Customers with questions should contact their local sales representative or Customer Service at 1-866-246-6990.

Details

Recall number
Z-0898-2017
Initiated
October 27, 2016
Posted by FDA
November 9, 2016
Terminated
September 25, 2020
Product code
CCQ – Nebulizer, Medicinal, Non-Ventilatory (Atomizer)
Quantity
2,674,650 units in total
Distribution
Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
Recalling firm
Teleflex Medical, Research Triangle Park, NC

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0381-2017 Class IILMA MADomizer, Rx only, Product Usage: These products are used for the delivery of…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated
Z-0382-2017 Class IIFLEXI NOZZLE LONG, 10 Boxes of 25 units each Product Usage: These products are used for…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated
Z-0383-2017 Class IIFLEXI NOZZLE SHORT Product Usage: These products are used for the delivery of topical…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated
Z-0384-2017 Class IILMA MADgic, Atomization Device with 3mL Syringe, Rx only, Product Usage: These products…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated
Z-0385-2017 Class IILMA MADett, Endotracheal tube Mucosal Atomization Device, Rx only, Product Usage: These…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated
Z-0386-2017 Class IILMA MADdy, Rx only, Product Usage: These products are used for the delivery of topical…Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication. 10/27/2016Terminated