Recall Z-0901-2017

Medtronic Xomed, Inc. · initiated June 28, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NIM EMG Endotracheal Tube, 6.0 mm I.D. x 8.8 mm O.D. (27 FR), Reinforced, Sterile, Rx only. Individually packaged.

Codes / lots: All lots distributed since May 2012. Catalog Ref No. (reinforced standard tubes) 82-29306, 82-29307, & 82-29308; Catalog Ref No. (reinforced standard tubes, Japan) 8229306J, 8229307J, & 8229308J; Catalot Ref No. (reinforced contact tubes) 8229506, 8229507, & 8229508;

Reason for recall

Wire in tubing can become exposed, posing potential harm to the intubated patient.

FDA-determined cause: Use error

Action taken

Medtronic sent an Urgent Product Safety Advisory letter dated June 28, 2016, to all affected customers informing them of the correction and providing instruction to avoid causing injury when using the EMG Endotracheal Tube. Recommended Actions included: inspecting the product thoroughly before use, testing the cuff by inflating it then deflating it before intubation, using a non-malleable stylet that closely matches the natural curve of the tube, not bending and flexing the tube excessively before use, and lubricating the cuff with a non-paralyzing, aqueous lubricant and a lubricated stylet. If customers found an EMG tube that appeared to have an electrode wire protruding from the distal end, the tube is not to be used and practitioners are urged to call Medtronic Product Quality Experience Department at 1-866-849-4003. For further questions , please call (904) 296-9600.

Details

Recall number
Z-0901-2017
Initiated
June 28, 2016
Terminated
August 7, 2019
Product code
ETN – Stimulator, Nerve
Quantity
3752
Distribution
Worldwide Distribution - Nationwide Distribution including US Virgin Island and Puerto Rico , and to the countries of :Japan, Korea, China, Canada, Dominican Republic, Colombia, Chile, Costa Rica, Bahamas, Peru, Panama, Mexico, Argentina, Ecuador, Brazil, Trinidad and Tobago, Hong Kong, Taiwan, United Kingdom, Netherlands, Sri Lanka, Austria, Italy, Germany, United Arab Emirates, Saudi Arabia…
Recalling firm
Medtronic Xomed, Inc., Jacksonville, FL

Official FDA recall record