Recall Z-0904-2025

ev3 Inc · initiated December 10, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090, Size: 5MM X 40MM X 90CM; 4) AB14W050040150, Size: 5MM X 40MM X 150CM; 5) AB14W050060150, Size: 5MM X 60MM X 150CM; 6) AB14W050080090, Size: 5MM X 80MM X 90CM; 7) AB14W050080150, Size: 5MM X 80MM X 150CM; 8) AB14W050100150, Size: 5MM X 100MM X 150CM; 9) AB14W050120150, Size: 5MM X 120MM X 150CM; 10) AB14W050150150, Size: 5MM X 150MM X 150CM; 11) AB14W050200090, Size: 5MM X 200MM X 90CM; 12) AB14W050200150, Size: 5MM X 200MM X 150CM; 13) AB14W060040090, Size: 6MM X 40MM X 90CM; 14) AB14W060040150, Size: 6MM X 40MM X 150CM; 15) AB14W060060150, Size: 6MM X 60MM X 150CM; 16) AB14W060080150, Size: 6MM X 80MM X 150CM; 17) AB14W060100150, Size: 6MM X 100MM X 150CM; 18) AB14W060120150, Size: 6MM X 120MM X 150CM; 19) AB14W060150150, Size: 6MM X 150MM X 150CM; 20) AB14W060200150, Size: 6MM X 200MM X 150CM; Balloon Dilatation Catheter

Codes / lots: 1) AB14W050020090, UDI/DI 00763000758035, Serial Numbers: B738596, B749440, B764968; 2) AB14W050020150, UDI/DI 00763000758042, Serial Numbers: B761521, B781264, B794814; 3) AB14W050040090, UDI/DI 00763000758059, Serial Numbers: B740725, B748628; 3) AB14W050040090, UDI/DI 04987742337891, Serial Numbers: B740725; 4) AB14W050040150, UDI/DI 00763000758066, Serial Numbers: B748644, B755679, B757663, B760286, B761769, B762343, B775924, B775972, B776408, B792995; 4) AB14W050040150, UDI/DI 04987742337907, Serial Numbers: B755679; 5) AB14W050060150, UDI/DI 00763000758080, Serial Numbers: B755683, B761779, B762646, B792990; 6) AB14W050080090, UDI/DI 00763000758097, Serial Numbers: B797860; 7) AB14W050080150, UDI/DI 00763000758103, Serial Numbers: B734091, B750081, B755684, B761172, B761754, B762256, B775959, B793874; 8) AB14W050100150, UDI/DI none, Serial Numbers: B750054, B758860, B761368, B764977, B765527, B769645, B781316, B791878, B805118; 9) AB14W050120150, UDI/DI 00763000758141, Serial Numbers: B731160, B761782, B762036, B768737, B793229; 10) AB14W050150150, UDI/DI 00763000758165, Serial Numbers: B727462, B736200, B750625, B758294, B758866, B760303, B762032, B763323, B763333…

Reason for recall

There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.

FDA-determined cause: Process change control

Action taken

Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 12/10/2024 via mail courier. The notice explained the issue, potential harms, and requested the following actions be taken: "" Immediately locate and quarantine all affected and unused NanoCross devices as listed in the provided scope. " Return all unused affected products in your inventory to Medtronic. Contact Medtronic Vascular Customer Service at 800-716-6700 to initiate a product return. Your local Medtronic Representative can assist you with the initiation of the return. " Complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and return it to RS.CFQFCA@medtronic.com within 30-days of receipt of this letter. " Please forward this notice to all those who need to be aware within your organization and maintain a copy for your records."

Details

Recall number
Z-0904-2025
Initiated
December 10, 2024
Posted by FDA
January 15, 2025
Product code
LIT – Catheter, Angioplasty, Peripheral, Transluminal
Quantity
5189 units
Distribution
Worldwide distribution - US (& PR) Nationwide and the countries of Chile, Cyprus, Estonia, Germany, Italy, Japan, Jordan, Puerto Rico, Spain, United States, Uzbekistan, and Yemen.
Recalling firm
ev3 Inc, Plymouth, MN

Official FDA recall record