Recall Z-0912-2018

Ortho-Clinical Diagnostics · initiated September 22, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS Chemistry Products DAT Performance Verifier II in vitro diagnostic

Codes / lots: Catalog# 6802328 Lot# J5617

Reason for recall

This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.

FDA-determined cause: Under Investigation by firm

Action taken

On September 18, 2017 a letter was sent to customers who received the BENZ reagent within the past 18 months of affected lots. The letter stated that customers may continue use of the product if QC results using VITROS DAT Performance Verifiers are within established ranges for each assay. If QC BENZ results are below established ranges, the customer should discontinue use of the product. The letter requests customers post the notification by each system that processes the VITROS BENZ Reagent. If the product was transferred outside of the facility, to notify those customers as well. Affected lots will be replaced in 6-8 weeks. Questions or concerns can be directed to 585-453-3452

Details

Recall number
Z-0912-2018
Initiated
September 22, 2017
Terminated
August 24, 2020
Product code
DIF – Drug Mixture Control Materials
Quantity
323
Distribution
Worldwide distribution: US Nationwide (including Puerto Rico), Bermuda, Australia, Brazil, Canada, Chile, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain, Poland, and Russia.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0911-2018 Class IIVITROS Chemistry Products DAT Performance Verifier I in vitro diagnosticThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers… 09/22/2017Terminated
Z-0913-2018 Class IIVITROS Chemistry Products DAT Performance Verifier III in vitro diagnosticThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers… 09/22/2017Terminated
Z-0914-2018 Class IIVITROS Chemistry Products DAT Performance Verifier IV in vitro diagnosticThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers… 09/22/2017Terminated
Z-0915-2018 Class IIVITROS Chemistry Products DAT Performance Verifier V in vitro diagnosticThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers… 09/22/2017Terminated