Recall Z-0919-2023
The Anspach Effort, Inc. · initiated December 7, 2022
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
Codes / lots: GTIN: 00845384001744 All lots
Reason for recall
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
FDA-determined cause: Under Investigation by firm
Action taken
DePuy Synthes initiated "URGENT: MEDICAL DEVICE CORRECTION (Notification)" via UPS on 12/07/22. Letter states reason for recall, health risk an action to be taken. The customers were instructed to do the following: Review your inventory and make a plan to follow the suggested manufacturer inspection interval of 12 months associated with your product, as specified in the IFU. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Once this notification is received, please complete, sign, and return the attached Business Response Form (page 3 of this letter) via Fax: 877-496-5042 or Scan/email: Synthes3606@sedgwick.com within 3 business days of receipt of this notification. Contact your Product Support Team through email at ra-dpyus-prodsupport@its.jnj.com or via phone at 1-800.327.6887, option #2 for additional support as needed.
Details
- Recall number
- Z-0919-2023
- Initiated
- December 7, 2022
- Posted by FDA
- January 7, 2023
- Product code
- HBC – Motor, Drill, Electric
- Quantity
- 6059 units in total
- Distribution
- US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania…
- Recalling firm
- The Anspach Effort, Inc., Palm Beach Gardens, FL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0916-2023 Class II | 6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems.…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0917-2023 Class II | 6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0918-2023 Class II | Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0920-2023 Class II | 7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref:…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0921-2023 Class II | Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0922-2023 Class II | 7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0923-2023 Class II | 6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref:…Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |
| Z-0924-2023 Class II | Pediatric Craniotome Ref:CRANI-P-G1Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment… | 12/07/2022Open, Classified |