Recall Z-0922-2026

Medline Industries, LP · initiated November 13, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

FORCEP, STD STR, STERILE KELLY, 5.5 IN, Item Number DYNJ04048

Codes / lots: DYNJ04048, UDI-DI: 10193489195248(each), 20193489195245(case), Lot Number: 2023112290

Reason for recall

The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.

FDA-determined cause: Packaging process control

Action taken

An URGENT MEDICAL DEVICE RECALL notification letter dated 11/13/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Your account will receive credit once the response form is submitted, if applicable. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-0922-2026
Initiated
November 13, 2025
Posted by FDA
December 12, 2025
Product code
LRW – Scissors, General, Surgical
Quantity
53 units
Distribution
US Nationwide distribution in the states CA, CO, IL, KS, and MI.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0921-2026 Class IISCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may… 11/13/2025Open, Classified