Recall Z-0928-2015

Trumpf Medical Systems, Inc. · initiated December 3, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Cable remote control JUC, BUTTON, Material no.1239823, Serial number: to (incl.) 100143048. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Codes / lots: The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

FDA-determined cause: Component change control

Action taken

Trumpf sent an Urgent Medical Device Correction letter on December 3, 2014, to all affected customers. Customers were instructed to complete and return the Confirmation of Receipt form with serial numbers of the affected table(s) and the serial number(s) for any remote controls they have. Replacement remotes will be sent. Customers were instructed to dispose of their recalled remotes. Customers with questions were instructed to call 888-474-9359. For questions regarding this recall call 843-534-0606.

Details

Recall number
Z-0928-2015
Initiated
December 3, 2014
Posted by FDA
January 6, 2015
Terminated
August 23, 2016
Product code
JEA – Table, Surgical With Orthopedic Accessories, Ac-Powered
Quantity
3303 total all devices
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
Recalling firm
Trumpf Medical Systems, Inc., Charleston, SC

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0918-2015 Class IICable remote control JUPITER U, Material no.4500119, Serial number: to (incl.)…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0919-2015 Class IICable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.)…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0920-2015 Class IICable remote control JUPITER, Material no. 4544119, Serial number: to (incl.) 102284378.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0921-2015 Class IIIR remote control JUPITER U, Material no. 4500120, Serial number: to (incl.) 102306940.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0922-2015 Class IIIR remote control JUPITER / TS7500, Material no.1601927, Serial number: to (incl.)…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0923-2015 Class IIIIR remote control JUPITER, Material no.4544120, Serial number: to (incl.) 102006720.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0924-2015 Class II(R)IR remote control JUPITER, Material no.1219656, Serial number: to (incl.) 102317869.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0925-2015 Class II(R)Cable remote control JUPITER, Material no.1219655, Serial number: to (incl.)…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0926-2015 Class IICable remote control JUX Material, no.1307315, Serial number: to (incl.) 101287649.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0927-2015 Class IIIR remote control JUX, Material no.1307314, Serial number: to (incl.) 102338310 Patient…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated
Z-0929-2015 Class IIIR remote control JUC, BUTTON, Material no.1239824, Serial number: to (incl.) 100139854.…Complaints have been reported of unintended movement occurring during surgical cases due to the use of… 12/03/2014Terminated