Recall Z-0928-2024
AtriCure, Inc. · initiated November 28, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056
Codes / lots: UDI-DI: 10840143903076; Lot Number: 129675
Reason for recall
AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
FDA-determined cause: Other
Action taken
An URGENT: MEDICAL DEVICE RECALL notification letter dated 11/28/23 was sent to customers. Actions to be taken by the Customer/User: " If the Tank Hose assembly is part of the lots listed below, do not use the tank hose assembly. " If the lot number is listed below, work with your AtriCure sales rep to replace the device with a new Tank Hose assembly and return the recalled device. " Return the attached Medical Device Recall Return Response Acknowledgement and Receipt Form. Type of Action by the Company: AtriCure is arranging for the removal and replacement of the affected tank hose assemblies. If you have impacted product, AtriCure will provide a replacement product at no charge to you. If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-755-4100) from 9-5pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for prompt review.
Details
- Recall number
- Z-0928-2024
- Initiated
- November 28, 2023
- Posted by FDA
- January 31, 2024
- Product code
- GEH – Unit, Cryosurgical, Accessories
- Quantity
- 1
- Distribution
- US Nationwide. Brazil, Canada, Japan, Taiwan
- Recalling firm
- AtriCure, Inc., Mason, OH
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0927-2024 Class II | Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect… | 11/28/2023Open, Classified |