Recall Z-0933-2021
Smith & Nephew, Inc. · initiated December 18, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.
Codes / lots: Product Number: 71271440; Batch Number: 18LC07540
Reason for recall
Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
FDA-determined cause: Storage
Action taken
A Recall notification letter titled, "Urgent Medical Device Recall Notice" was sent to consignees on 12/18/2020 via email and FedEx. The letter instructs the consignee to follow the instructions on the attached response form. The response form provides the following instructions: "1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com."
Details
- Recall number
- Z-0933-2021
- Initiated
- December 18, 2020
- Terminated
- May 1, 2023
- Product code
- LOD – Bone Cement
- Quantity
- 5190 devices
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.
- Recalling firm
- Smith & Nephew, Inc., Memphis, TN