Recall Z-0933-2022

MEDLINE INDUSTRIES, LP - Northfield · initiated March 8, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stanbio TDM/Beta-Hydroxybutyrate Linearity Standards, packaged in boxes containing Levels 1-6, 1 x 4 mL each, REF 2450-604, Medline catalog number S-O2450604, IVD.

Codes / lots: Lot numbers 172093, exp. 06/30/2023; 174703, exp. 08/31/2023; and 169141, exp. 04/30/2023.

Reason for recall

The product was stored incorrectly due to improper storage controls which may lead to delayed results.

FDA-determined cause: Storage

Action taken

The recalling firm issued letters dated 3/8/2022 via regular mail on 3/8/2022. The letter explained the product was stored incorrectly due to improper storage controls. The consignee was to immediately check their inventory for the affected lot numbers, quarantine it, and destroy it. The enclosed destruction form was to be completed listing the quantity of affected product destroyed and once the form was returned, they would receive credit. If the product was transferred to another individual or location, they were to notify them of the recall communication. On 3/17/2022, the recalling firm issued an amended recall letter dated 3/17/2022 that notified the consignee there was an error in one of the lot numbers (1747003) listed in the original letter dated 3/8/2022 and the correct lot number is 174703. The other lot numbers were also listed in the amended letter. Again, return of the amended Destruction Form listing all of the lot numbers was requested containing the same instructions as the original Destruction Form.

Details

Recall number
Z-0933-2022
Initiated
March 8, 2022
Terminated
February 27, 2023
Product code
DKB – Calibrators, Drug Mixture
Quantity
10 boxes
Distribution
Distribution was made to AL, FL, NE OH, TX, WA, and WI. There was no foreign/military/government distribution.
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record