Recall Z-0938-2026

Instrumentation Laboratory · initiated November 18, 2025

Class III Open, Classified

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

HemosIL SynthAFax. Partial Thromboplastin Time Tests.

Codes / lots: Part No. 0020007400; UDI: 08426950087649; Lot No. (Exp. Date) N0542231(2026-06-30), N0250088(2027-04-30, N0452249 (2027-06-30).

Reason for recall

Recalled lots were manufactured with double the amount of preservative concentration.

FDA-determined cause: Process control

Action taken

An URGENT MEDICAL DEVICE RECALL notification, dated 11/18/25, was sent to consignees. Consignees are to contact Client Services at 1-800-955-9525 to arrange conversion to replacement lot N0956748, and in the meantime continue use of affected lots with the QC lot reassignments provided and run QC per labeled instructions for use. Once the replacement lot is received, consignees are to complete the cross-over study per laboratory procedures and immediately destroy affected devices, documenting destruction on the provided Mandatory Urgent Medical Device Recall Tracking Form which should be returned to WERFEN by fax at 781-861-4207 or by email to ra-usa.na@werfen.com. All laboratory staff are to be notified of this recall and a copy of the notice posted near IL Coagulation Systems. Consignees with any questions are to contact Client Services at 1-800-955-9525.

Details

Recall number
Z-0938-2026
Initiated
November 18, 2025
Posted by FDA
December 17, 2025
Product code
GFO – Activated Partial Thromboplastin
Quantity
4,506 units
Distribution
Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
Recalling firm
Instrumentation Laboratory, Bedford, MA