Recall Z-0953-2017

In2Bones, SAS · initiated December 19, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NEO SCREW 02.7-14MM LOCKING, REF W27 ST114, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW Plating System is intended for fixation of intra-articular and extraarticular fractures of the distal radius and reconstruction of the distal radius

Codes / lots: Lot Numbers: 1604154, 1605083, 1606068

Reason for recall

Dimensional non-conformity of the thread on the screw head.

FDA-determined cause: Component design/selection

Action taken

IN2BONES en a Field Safety Corrective Action letter dated December 15, 2016, to the sole US distributor. The letter identified the product, the problem, and the action to be taken by the customer. We therefore recommend you to follow the instructions here below: 1. Identify all the devices with the batches subject of this recall that might still be in your inventory and quarantine them 2. Inform and distribute this Recall Notification to all relevant persons within your organization 3. To distributors only: Identify all the devices with the batches subject of this recall that were delivered to your customers, and if relevant, instruct them to also follow these instructions (identification and quarantine) 4. Fill in and return the fax back form enclosed. With this form, you will certify that you have received this Recall Notification and intend to comply with the recommendations listed. This fax back form will enable In2Bones to conduct effectiveness checks. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at +33 4 72 29 26 26 / +33 7 61 88 44 30 or by email: qualte@in2bones.com. For questions regarding this recall call 844-602-6637.

Details

Recall number
Z-0953-2017
Initiated
December 19, 2016
Posted by FDA
January 6, 2017
Terminated
February 1, 2017
Product code
HRS – Plate, Fixation, Bone
Quantity
40 units
Distribution
Worldwide Distribution - US TN only and Internationally to France and Switzerland
Recalling firm
In2Bones, SAS, Ecully

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0952-2017 Class IISCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO, In2Bones, 28 chemin du petit…Dimensional non-conformity of the thread on the screw head. 12/19/2016Terminated
Z-0954-2017 Class IINEOVIEW(R) SCREWS SET, REF W27 ST900, STERILE R The NEOVIEW Plating System is intended…Dimensional non-conformity of the thread on the screw head. 12/19/2016Terminated