Recall Z-0957-2024

Wipro GE Healthcare Private Ltd. · initiated December 26, 2023

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000

Codes / lots: a) Care Plus, Model Numbers: 6600-0040-900, 6600-0203-900, 6600-0203-903, 6600-0232-910, 6600-0232-920, 6600-0235-901, CARE-PLUS; b) Care Plus 1000, Model Numbers: 8200, 6600-0264-901, 6600-0264-903, 6600-0265-901, 6600-0265-902, 6600-0265-903, 6600-0265-908, 6600-0266-901, 6600-0266-903, 6600-0267-901, 6600-0269-901, 6600-0269-903, 6600-0270-901, 6600-0272-901; c) Care Plus 2000, Model Numbers: 2073175-001, 6600-0273-901, 6600-0274-901, 6600-0274-903, 6600-0274-911, 6600-0275-901, 6600-0275-902, 6600-0275-903, 6600-0275-908, 6600-0275-911, 6600-0276-901, 6600-0276-903, 6600-0277-901, 6600-0279-903, 6600-0280-901, 6600-0280-903, 6600-0281-903, 6600-0282-901, M1050218, M1160968, M1206249, M1242714, QCR043; d) Care Plus 3000, Model Numbers: 6600-0203-902, 6600-0223-981, 6600-0230-900, 6600-0230-901, 6600-0232-901, 6600-0232-911, 6600-0232-921, 6600-0232-980, 6600-0232-983, 6600-0233-900, 6600-0233-901, 6600-0233-911, 6600-0235-900, 6600-0293-901, 6600-0293-903, 6600-0294-901, 6600-0294-902, 6600-0294-903, 6600-0294-905, 6600-0294-908, 6600-0294-909, 6600-0295-901, 6600-0301-901, 6600-0301-903, 6600-0302-901, 6600-0302-903, 6600-0303-901, 6600-0333-903, 6600-0463-901, 6600-0511-901…

Reason for recall

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

FDA-determined cause: Labeling design

Action taken

GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/26/2023 by letter using a traceable means. The notice explained the issued, attached copies of the End of Sales/Service Support letters that were previously distributed, and recommended the replacement of the devices, if possible. Customers who wish to continue using these devices should ensure that: " Each time the bedside panel is closed, the latches are secure. " Each time the porthole doors are closed, the porthole latches are manually checked to make sure the porthole door latches are secure. " Each time the bedside panel or pothole is closed, the porthole door is pulled to make sure it is latched. GE HealthCare will provide a user manual addendum, and label for the device, in a future mailing at no cost to you. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Details

Recall number
Z-0957-2024
Initiated
December 26, 2023
Posted by FDA
February 9, 2024
Product code
FMZ – Incubator, Neonatal
Quantity
21626 units
Distribution
Worldwide distribution.
Recalling firm
Wipro GE Healthcare Private Ltd., Bengaluru

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0958-2024 Class IGE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP…GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and… 12/26/2023Open, Classified
Z-0959-2024 Class IGE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers:…GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and… 12/26/2023Open, Classified
Z-0960-2024 Class IGE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800…GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and… 12/26/2023Open, Classified
Z-0961-2024 Class IGE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659…GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and… 12/26/2023Open, Classified
Z-0962-2024 Class IGE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767…GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and… 12/26/2023Open, Classified