Recall Z-0960-2026

Cepheid · initiated October 30, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-10

Codes / lots: UDI: 007332940001834/Lot Numbers: 1001452775 1001457465 1001457622 1001459091 1001462586 1001462625 1001463944 1001465391 1001469480 1001473229 1001474930 1001477474 1001480636 1001483324 1001486594 1001492908 1001493823 1001495811 1001498202 1001501650

Reason for recall

Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.

FDA-determined cause: Process design

Action taken

On November 14, 2025, Cepheid issued a "Urgent Medical Device Recall" notification to affected consignees. Cepheid asked consignees to take the following actions: 1. Cepheid recommends that laboratories and clinical facilities utilizing the Xpert BCR-ABL Ultra assay for quantification of BCR::ABL1 and ABL1 mRNA transcript levels in peripheral blood specimens from t(9;22)-positive chronic myeloid leukemia (CML) patients undergoing tyrosine kinase inhibitor (TKI) therapy, assess white blood cell count (WBCC) prior to testing in cases where elevated white blood cell count is suspected. 2. If WBCCC exceeds 30,000 cells/uL, the specimen should be processed in the accordance with the Retest Procedure for ERROR (Code 2008) or INVALID (Type 2) as detailed in the IFU for the following scenarios: - Cases where INVALID test results are observed and are associated with an elevated WBCC. - Cases of valid results in which an elevated WBCC has been observed. 3. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 4. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them with a copy of this letter. 5. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication.

Details

Recall number
Z-0960-2026
Initiated
October 30, 2025
Posted by FDA
December 18, 2025
Product code
OYX – Bcr/Abl1 Monitoring Test
Quantity
4032 units
Distribution
US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe…
Recalling firm
Cepheid, Sunnyvale, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0959-2026 Class IIBrand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may… 10/30/2025Open, Classified