Recall Z-0965-2017

Advanced Sterilization Products · initiated September 27, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

EVOTECH Endoscope Cleaner and Reprocessor System 208V, Product Code: 50004 The EVOTECH ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX OPA Concentrate Solution and an enzymatic detergent (CIDEZYME XTRA) to achieve cleaning and high level disinfection of heat sensitive (>60C} semi-critical endoscopes.

Codes / lots: UDI 10705037030958 Serial No. 5041150187 UDI 10705037030958 Serial No. 5041150188 UDI 10705037030958 Serial No. 5041150189 UDI 10705037030958 Serial No. 5041150192 UDI 10705037030958 Serial No. 5041150191 UDI 10705037030958 Serial No. 5041150190 UDI 10705037030958 Serial No. 5041150193 UDI 10705037030958 Serial No. 5041150205 UDI 10705037030958 Serial No. 5041150212 UDI 10705037030958 Serial No. 5041150207 UDI 10705037030958 Serial No. 5041150213 UDI 10705037030958 Serial No. 5041150206 UDI 10705037030958 Serial No. 5041150204 UDI 10705037030958 Serial No. 5041150210 UDI 10705037030958 Serial No. 5041150198 UDI 10705037030958 Serial No. 5041150228 UDI 10705037030958 Serial No. 5041150230 UDI 10705037030958 Serial No. 5041150229 UDI 10705037030958 Serial No. 5041150227 UDI 10705037030958 Serial No. 5041150226 UDI 10705037030958 Serial No. 5041150209 UDI 10705037030958 Serial No. 5041150231 UDI 10705037030958 Serial No. 5041150197 UDI 10705037030958 Serial No. 5041150236 UDI 10705037030958 Serial No. 5041150211 UDI 10705037030958 Serial No. 5041150235 UDI 10705037030958 Serial No. 5041150233 UDI 10705037030958 Serial No. 5041150234 UDI 10705037030958 Serial No. 5041150232 UDI…

Reason for recall

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.

FDA-determined cause: Nonconforming Material/Component

Action taken

ASP has sent out Service Representatives to all impacted customers to inspect the hoses for the EVOTECH Endoscope Cleaner and Reprocessor System and will replace the hose if required. The firm issued a Technical Bulletin dated 8/22/2016. Customers were instructed to complete online training prior to performing the actions described in the bulletin and follow instructions. For orders and questions call 888-783-7723 or email: RA-ASPUS-ServPartOrd@aspus.jnj.com.

Details

Recall number
Z-0965-2017
Initiated
September 27, 2016
Posted by FDA
January 9, 2017
Terminated
April 17, 2017
Product code
FEB – Accessories, Cleaning, For Endoscope
Quantity
107 units
Distribution
US Nationwide Distribution.
Recalling firm
Advanced Sterilization Products, Irvine, CA

Official FDA recall record