Recall Z-0967-2013

Alphatec Spine, Inc. · initiated July 6, 2011

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ILLICO Straight CP TI Rod, -5.5MM x 110M; REF#73601-110; Qty 1; LBL-005; REV.B; RX ONLY; Alphatec Spine, Inc. The ILLICO MIS Posterior Spinal Fixation System is intended to facilitate the surgical correction of non-cervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.

Codes / lots: Lot # 635436; Part #73601-110

Reason for recall

The rods were incorrectly laser etched as ILLICO Straight CP Tl Rods, 110mm, when they were in fact manufactured to 100mm in length.

FDA-determined cause: Process control

Action taken

Alphatec Spine sent an Urgent Medical Device Recall letter dated July 6, 2011, to all affected customers. This notification described the product, problem and actions to be taken. The consignees were instructed to review their inventory to verify the affected device; abstain from use; immediately contact Alphatec Spine Sales Administration (salesadministration@alphatecspine.com) for instructions on how to return this affected product; and complete, sign and return confirmation form to Alphatec Spine, Inc., 5818 El Camino Real, Carlsbad, CA 92008. If you have any questions, call 760-431-9286.

Details

Recall number
Z-0967-2013
Initiated
July 6, 2011
Posted by FDA
March 18, 2013
Terminated
March 20, 2013
Product code
KWP – Appliance, Fixation, Spinal Interlaminal
Quantity
7
Distribution
Worldwide Distribution - USA including CO and NC; Internationally to Belgium.
Recalling firm
Alphatec Spine, Inc., Carlsbad, CA

Official FDA recall record