Recall Z-0988-2011

Southmedic, Inc. · initiated September 3, 2010

Terminated

Product

Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0610, Size #10, General Surgical Use, Sterile, unless package is damaged or opened., Southmedic Inc., 50 Alliance Blvd., Barrie, ON., Canada L4M 5K3 www.southmedic.com, Made in Canada.

Codes / lots: Lot numbers: 070610, 82010, 82610, and 91710.

Reason for recall

Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile.

FDA-determined cause: Packaging process control

Action taken

On 9/3/2010, the firm notified consignees about the recall by facsimile letter. The letter instructed consignees to return scalpels for credit or replacement. The firm enlarged the scope of the recall on 11/15/2010. An additional 927 cases were recalled. This includes 11 additional lots. The firm sent out URGENT MEDICAL DEVICE RECALL letters dated 11/15/2010 to its direct accounts. The letter identified that affected product and the reason for the recall. The consignees were instructed to check their inventory, complete and return the attached Voluntary Recall Notification Confirmation Form, return the affected product to the manufacturer. Product return question should be directed towards Marc Pauze at 800-463-7146 ext 299 and regulatory questions should be directed towards Tish Anger at 800-463-7146 ext 342.

Details

Recall number
Z-0988-2011
Initiated
September 3, 2010
Posted by FDA
January 31, 2011
Terminated
May 18, 2011
Product code
GDX – Scalpel, One-Piece
Quantity
1330 all products
Distribution
Worldwide Distribution -- USA, including states of TX, MO, NJ, LA, IL, NY, KY, PA, VA, WA, CA, MA, MD, IN, NV, TN, NC, AZ, IA, WI, UT, NE, GA, FL, MN, ME, OH, SD, and KS and countries of Australia and Canada.
Recalling firm
Southmedic, Inc., Barrie Canada, Ontario

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0989-2011 Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0611, Size…Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile. 09/03/2010Terminated
Z-0990-2011 Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0615, Size…Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile. 09/03/2010Terminated