Recall Z-0992-2023
Cardinal Health 200, LLC · initiated December 14, 2022
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Cardinal Health Jackson-Pratt Channel Drain, 24 Fr, REF JP-2234
Codes / lots: UDI/DI 50885380182660(cs), 20885380182669(ea), Lot Numbers: N210278, N210347, N210368, N210423, N210460, N210468, N210469, N210543, N220050, N220099, N220139, N220163
Reason for recall
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
FDA-determined cause: Unknown/Undetermined by firm
Action taken
Cardinal Health issued an URGENT MEDICAL DEVICE PRODUCT RECALL notice to its consignees on 12/14/2022 by overnight mail. The notice explained the issue, the risk, and requested the following: Actions Required: 1. REVIEW your inventory for the affected product codes and lots. Location of product codes and lots are shown above. 2. SEGREGATE and quarantine all affected product upon review of your inventory. Product should not be used and cease using the product immediately. 3. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product recall notice and share a copy of this notice. 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMBFieldCorrectiveAction@cardinalhealth.com, whether you have affected product. Return of Product and Available Assistance: CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5pm EST: . Hospital 800-964-5227 . Federal Government 800-444-1166 . Distributor 800-635-6021 . All other customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Details
- Recall number
- Z-0992-2023
- Initiated
- December 14, 2022
- Posted by FDA
- January 18, 2023
- Product code
- GBX – Catheter, Irrigation
- Quantity
- 22847 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.
- Recalling firm
- Cardinal Health 200, LLC, Waukegan, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0967-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2186Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0968-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2187Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0969-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2188Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0970-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2189Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0971-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2190Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0972-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2191Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0973-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2210Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0974-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2211Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0975-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2212Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0976-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2213Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0977-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2214Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0978-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2215Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0979-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2216Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0980-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2217Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0981-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2221Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0982-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2223Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0983-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2225Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0984-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2226Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0985-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2227Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0986-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2228Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0987-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2229Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0988-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2230Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0989-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2231Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0990-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2232Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0991-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2233Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0993-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 28 Fr, REF JP-2290Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |
| Z-0994-2023 Class II | Cardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal… | 12/14/2022Open, Classified |