Recall Z-0997-2020

Roche Diabetes Care, Inc. · initiated December 11, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Accu-Chek Aviva Plus Health Network Strips- intended for Blood Glucose Monitoring System Catalog Number: 06908349001

Codes / lots: Lot Number: 497864 NDC: 65702-0438-10 UPC: 3-65702-43810-1

Reason for recall

Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions

FDA-determined cause: Process control

Action taken

Roche Diabetes Care issued Urgent Medical Device Corrections via UPS Ground mail to the patients and distributors on 12/11/2019. Letter states reason for recall, health risk and action to take: Immediately discontinue using test strip lots 497864 and 498315. Affected test strips must be discarded. To get replacement test strips, or for questions regarding this notification, use one of the following options: For fastest replacement of your product, please submit your request at https://notices.accu-chek.com. Email Accu-Chek Customer Care at accu-chek.care@roche.com and include the following information: your name, full mailing address, phone number, test strip lot number and number of affected test strip vials. o Contact our Accu-Chek Customer Care line at 1-800-358-4866 Monday through Friday between the hours of 8:00 am and 8:00 pm Eastern Standard Time. If a Business Reply Form was included with this notification, please complete and return it by U.S. mail using the provided postage paid envelope. Please keep this notification for future reference. Distributors/Suppliers instructed as follows: Discontinue distribution and destroy any remaining stock of the affected lots (497864 and 498315) immediately. make this notification available to consumer customers to whom you distributed affected product. If you have distributed affected product to other suppliers, provide a copy of this UMDC to those suppliers. Have those customers report the quantity of product destroyed to you.

Details

Recall number
Z-0997-2020
Initiated
December 11, 2019
Terminated
July 8, 2022
Product code
NBW – System, Test, Blood Glucose, Over The Counter
Quantity
82,332 vials
Distribution
Nationwide: Foreign: Austalia, India
Recalling firm
Roche Diabetes Care, Inc., Indianapolis, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0998-2020 Class IIAccu-Chek Aviva Plus Retail Strips 50 ct - intended for Blood Glucose Monitoring System…Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a… 12/11/2019Terminated
Z-0999-2020 Class IIACCU-CHEK AVIVA ASPAC STRIP 50CT- intended for Blood Glucose Monitoring System Catalog…Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a… 12/11/2019Terminated