Recall Z-1001-2020

Clinical Diagnostic Solutions, Inc. · initiated January 2, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011

Codes / lots: Part Number 1400075 (US) Part Number 1400011 (OUS) All serial numbers distributed prior to 3/21/2019

Reason for recall

A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.

FDA-determined cause: Software design

Action taken

Consignees notified via email and postal delivery with recall letter dated 01/02/2020. The letter identified affected product, stated reason for recall, and provided specific actions to follow. A Recall Acknowledgement Form was requested to be returned by email, postal delivery, or fax. Clinical Diagnostic Solutions Inc. is handling US consignees only. Boule Medical AB is handling OUS consignees.

Details

Recall number
Z-1001-2020
Initiated
January 2, 2020
Terminated
November 3, 2022
Product code
GKZ – Counter, Differential Cell
Quantity
261
Distribution
Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
Recalling firm
Clinical Diagnostic Solutions, Inc., Plantation, FL