Recall Z-1003-2019

Ossur Americas · initiated June 21, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Ossur, ReSolve Halo Vest, Part/ Description: 545700K/ ReSolve Halo Vest-shrt kodel; 545700L/ ReSolve Halo Vest-shrt lmbwool; 545800K/ ReSolve Halo Vest-tall kodel; 545800L/ ReSolve Halo Vest-tall lmbwool

Codes / lots: All lots of the products manufactured from 06/30/2009 - 12/13/2018

Reason for recall

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

FDA-determined cause: Labeling design

Action taken

On 06/21/2018, Urgent Field Safety Notices were mailed to customers via UPS Next-Day Air requiring a signature of acceptance from the customer. Customers were advised the following: 1. Please pass this notice to those who need to be aware within your organization or to any department areas where the potentially affected devices will be used or transferred. 2. Maintain awareness on this notice for an appropriate period. 3. If you would like to have the product replaced please contact your customer service or sales representative and new product will be provided. Customers can exchange the affected product, if they desire to do so. Customers were asked to distribute the notices to those who need to be aware within their organization or to any department areas where the potentially affected devices were used or transferred. On 01/14/2019, Urgent Medical Device Recall letters were sent via express mail requiring a signature of acceptance from the customer. Customers were advised: Please examine your inventory, quarantine products subject to the recall and contact customer service for a return authorization to return the product. In addition: 1. Please confirm receipt of this recall notice by calling customer service (800-233-6263) or visit https://safetyalert.ossur.com 2. Please pass this notice to those who need to be aware within your organization and to any department areas where the potentially affected devices will be used or transferred (e.g. MR Imaging, Surgical, Emergency, Material Management, Patient) 3. Maintain awareness on this notice for an appropriate period. 4. If you have further distributed this product, please identify your customers and notify them at once of this product alert. We recommend that you include a copy of this product recall. 5. If any of your patients is currently wearing the product, necessary arrangements should be made to ensure the patient is not placed in MR Imaging.

Details

Recall number
Z-1003-2019
Initiated
June 21, 2018
Terminated
September 4, 2020
Product code
KQZ – Component, Traction, Non-Invasive
Quantity
381
Distribution
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
Recalling firm
Ossur Americas, Foothill Ranch, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0997-2019 Class IIOssur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-0998-2019 Class IIOssur Skull Pin, Sterile Ceramic Tip Pin, Part/ Description: 516CS/ Skull Pin 2-1/4in CT…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-0999-2019 Class IIOssur OB Resolve, Sterile, Part/ Description: 505300D/ OB RESOLVE STERILE RING SML…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-1000-2019 Class IIOssur CB Resolve, Sterile, Part/ Description: 505300C/ CB RESOLVE STERILE UNIT SML…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-1001-2019 Class IIOssur, Sterile, Part/ Description: 505300T/ CB RING KIT TI SM; 505300V2/ OB V2 RING KIT…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-1002-2019 Class IIOssur, J Tong, Sterile, Part/ Description: JT-115/ J TONG MED/LGE-No App; JT-200/ J TONG…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated
Z-1004-2019 Class IIOssur, LIL Angel Vest, Part/ Description: 545100K/ LIL ANGEL VEST P2 KODEL; 545100L/ LIL…Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR… 06/21/2018Terminated