Recall Z-1011-2012

Bio-Detek, Inc. · initiated December 30, 2011

Terminated

Product

Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators specifically for pediatric patients.

Codes / lots: Lot Numbers 0510 thru 4511

Reason for recall

Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.

FDA-determined cause: Process control

Action taken

Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.

Details

Recall number
Z-1011-2012
Initiated
December 30, 2011
Posted by FDA
February 13, 2012
Terminated
December 11, 2012
Quantity
3,080 units
Distribution
Worldwide Distribution -- US and Canada.
Recalling firm
Bio-Detek, Inc., Pawtucket, RI

Official FDA recall record