Recall Z-1012-2012

Cytophil, Inc · initiated January 11, 2012

Terminated

Product

Model #03001-02: Periophil Biphasic, Synthetic Biphasic Dental Bone Graft Granules, 1.0 cc. Model # 03003-01 Periobase Biphasic: Periobase, Synthetic Biphasic Dental Bone Biphasic Granules, 0.5 cc. Model #03002-01: Periobone Biphasic, Synthetic HA/TCP Dental Bone Graft Biphasic Granules, 0.5 cc. Periophil Biphasic (synthetic grafting material) is intended for use as a bone grafting material to fill, augment, or reconstruct periodontal or oral/maxillofacial defects. Cytophil, Inc., 2485 Corporate Circle, Suite 2, East Roy, WI 53120.

Codes / lots: 1. 03001-02 Periophil Biphasic, Lot # Z13-927A, expiration date 2/1/2014. 2. 03001-02 Periophil Biphasic, Lot # Z13-927B, expiration date 2/1/2014. 3. 03001-01 Periophil Biphasic, Lot # Z13-926A, expiration date 2/1/2014. 4. 03001-01 Periophil Biphasic, Lot # N13-925C, expiration date 1/31/2014. 5. 03003-01 Periobase Biphasic, Lot # N13-925B, expiration date 1/31/2014. 6. 03002-01 Periobone Biphasic, Lot # N13-925A, expiration date 1/31/2012.

Reason for recall

Cytophil Corporation is recalling Periophil Biphasic Periophil Biphasic, Periobase Biphasic, and Periobone Biphasic because devices were inadvertently shipped after receiving an inadequate dose of radiation and this product may not be sterile.

FDA-determined cause: Process control

Action taken

Consignees were sent an "Urgent Medical Device Recall" letter dated January 11, 2012 by Cytophil. The letter described the problem and the product involved in the recall. The firm advised consignees to immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, they were requested to identify their customers and notify them at once of this product recall. For questions, call 262-642-2765.

Details

Recall number
Z-1012-2012
Initiated
January 11, 2012
Posted by FDA
February 13, 2012
Terminated
May 1, 2012
Product code
LYC – Bone Grafting Material, Synthetic
Quantity
547
Distribution
Nationwide Distribution - Including the states of WI, GA, IL, SC, MO, and TX.
Recalling firm
Cytophil, Inc, East Troy, WI