Recall Z-1014-2015

Clinical Innovations, LLC · initiated December 30, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Ebb Complete Tamponade System, BD-OTS Obstetric Catheter, REF CTS-1000, Rx Only, Sterile EO

Codes / lots: Lots: 0314-F-566; 0814-F-100; 0814-F-108; 0814-V-068

Reason for recall

Potential for a balloon leak in all lots of (CTS-1000) manufactured since August 2014.

FDA-determined cause: Process design

Action taken

Clinical Innovations sent an "Urgent - Medical Device Recall" letter dated December 31, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They will be instructed to quarantine remaining product. A company representative will visit the hospital and retrieve the product in accordance to instructions detailed on the Representative Recall Reconciliation Instructions. The representative will reconcile the inventory on the Field Reconciliation Form and return the product to Clinical Innovations as indicated on the instructions. The address for the devices to be returned is included on the Representative Recall Reconciliation Instruct Clinical Innovations is working diligently to resolve this issue. In the meantime ,please quarantine any remaining product in your hospital. A company representative will contact you and make arrangements for the return of the product. Clinical Innovations will provide replacement product as soon as it becomes available. If you have any questions,call Clinical Innovations,at 1-(888)268-6222 or your service representative, as is the name of the Company Contact.

Details

Recall number
Z-1014-2015
Initiated
December 30, 2014
Posted by FDA
January 27, 2015
Terminated
December 28, 2015
Product code
OQY – Intrauterine Tamponade Balloon
Quantity
827 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of : Canada, Finland, Estonia, Belgium, Ireland, Sweden, France, Switzerland, Spain, Netherlands, Middle East, Scotland, Germany and Wales.
Recalling firm
Clinical Innovations, LLC, Murray, UT

Official FDA recall record